A13914AF
9-17
Maintenance
Check Chloride Calibration Span
9
Check Chloride Calibration Span
The system automatically monitors the CL span with each calibration. If the span falls below 1800,
the system highlights the
Check Chloride Calibration Span
task in the Weekly tab of the Electronic
Maintenance Log. When the task is highlighted, replace the chloride electrode tip as soon as
convenient. See
Replace the Chloride Electrode
.
To check the chloride calibration span manually:
1
During the daily calibration routine, the AQUA CAL 2 Calibration Report is automatically
printed. If necessary, reprint this report. See
.
2
Refer to page 2 of the calibration report. Check the CL span value at the bottom of the page. If
the value is <1800, replace the electrode tip when it is convenient to your work load. See
Monthly Maintenance
Adjust the Fluid Level of the Alkaline Buffer Damper Assembly
BUNm/UREAm Electrode Maintenance (DxC 800 Only)
Replace CTS Blade (1-Blade Narrow CTS Option)
Replace Alkaline Buffer
The following procedure describes how to replace the Alkaline Buffer reagent and the In-line air
filters.
1
Loosen the two Phillips screws on the ISE module cover and remove the cover.
2
Release the locking pin (1) on the ISE module and lift the module until it clicks into a raised,
locked position. The locking pin is located on the left side of the ISE module on DxC 800 Systems
as shown in
. The locking pin is located on the right side of the ISE module on DxC 600
Systems.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......