A13914AF
9-67
Maintenance
As-Needed/As-Required Maintenance
9
IMPORTANT
Do not scratch the electrode port.
8
Install the retainer nut and a new quad-ring on the electrode and insert the new electrode into
electrode port. Turn electrode retainer nut until finger-tight.
9
Gently pull on electrode body to test for proper seating of electrode. The electrode assembly
should not move. If electrode moves, remove it and install it again. Check for an extra or missing
quad-ring if installation is difficult.
10
Reposition flow cell on the mounting panel and tighten the one retaining screw. Reconnect
electrode cable to the appropriate terminal being careful not to bend the connecting pin.
11
Select
OK
in the Maintenance pop-up confirmation window when the electrode service has
been completed and the electrode has been properly replaced.
IMPORTANT
Reagents immediately prime through the flow cell.
While the system is priming, observe the flow cell for leaks. Stop the priming if any leaks are
noticed and correct the problem.
12
When the system has completed priming, select
Close
to exit the procedure.
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
13
Lower the ISE module by releasing the locking pin. Be careful not to pinch tubing as the ISE
module is lowered.
14
Reinstall the ISE module cover.
15
Calibrate all ISE chemistries. A reference drift for new electrodes may be observed during the
first 4–6 hours after electrode replacement. Results will
NOT
be compromised. Recalibrating the
chemistry during this time period helps stabilize the electrode.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......