A13914AF
5-8
Quality Control
Run Control Samples
Run Control Samples With Bar Code Labels and Auto Generation
of Control Enabled
1
Place bar-coded control samples in rack. All tube bar code labels must be visible through the
slots on the same side of the rack as the rack bar code label.
2
When Auto Generation of Control is enabled, the automatic Multiple Cartridge option is
available. With the Multiple Cartridge option enabled, the system automatically runs all
cartridges onboard and runnable for any chemistries for an auto generated control. To enable
Multiple cartridge from the
Setup
screen, scroll down and select
29 Auto Generation of Control
.
3
Press
RUN
on the analyzer.
The instrument automatically runs chemistries that are:
•
Defined for the control.
•
Loaded and calibrated.
•
Contain sufficient volume to run the controls.
4
If all tests programmed for an Auto Generated control ID do not complete testing (for example,
reagent not onboard, calibration required, cal time exceeded or disabled), the next time the
control ID is loaded, all tests for the control will Auto Generate and be run again. If a control ID
that was Manually programmed does not complete testing, the next time the control ID is
loaded, only the incomplete test will be run. A Manually programmed control ID that is
incomplete must be manually cleared before reloading the sample in order to Auto Generate all
tests for the control.
Manually Programming Control Samples
1
Select
Samples
from the menu bar.
2
Clear the QC sample ID to be used.
•
Select
Clear F7
.
•
Type the QC sample ID into the
Sample ID(s)
field.
•
Select
OK
.
•
Select
OK
to confirm.
3
Select
Control F5
.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......