A13914AF
9-18
Maintenance
Monthly Maintenance
Figure 9.7
DxC 800 ISE Module
3
Unscrew and remove the bottle cap from the alkaline buffer reagent. Hold the cap with attached
straw over a container that catchs the residual fluid from the alkaline buffer return line #33.
4
Prime the line 10 times.
•
Select:
—
Utils
from the menu bar,
—
1 Prime
,
—
MC F4
,
—
ISE CO
2
Alkaline Buffer
.
•
Type
10
in the
Number of primes to repeat
field.
•
Select
Start Prime
. Reagent line is drained of old reagent.
5
Remove reagent input line #30 (1) (line draining into damper) from the top of the in-line filter.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......