A13914AF
4-19
Reagent Load/Calibration
Within-Lot Calibration
4
•
Chem name
•
Reagent Position
MC Subsystem Calibration Failure
The MC Subsystem warns the operator of a failed calibration. If a chemistry fails, a pop-up message
appears after a completed calibrator level. The message includes:
•
Chem name
•
Calibration error
•
Calibrator Name and Level Number
IMPORTANT
Troubleshooting Calibration and Result Errors
for additional information
on Calibration Error messages.
Within-Lot Calibration
Introduction
The Within-Lot Calibration option provides multiple reagent load of cartridges of the same lot
number without having to calibrate each cartridge. The calibration factor established from a fresh
cartridge (less than 8 hours on board the instrument) is stored and applied to subsequently loaded
cartridges of the same lot.
Within-Lot Calibration Frequency
The length of time a within-lot calibration factor remains valid (its within-lot calibration
frequency) depends on the chemistry. During this interval any newly loaded cartridge with the
same lot number will receive the calibration status
Calibrated
. At the end of this frequency period,
a new within-lot calibration factor must be established from a fresh cartridge.
The calibration frequency of a particular cartridge has priority over the within-lot calibration
frequency. If a cartridge selected for within-lot calibration is still loaded at the end of its individual
calibration frequency period, that cartridge must be recalibrated (this will not affect the within-lot
calibration factor).
Enable Within-Lot Calibration
The system status must be
Standby
.
1
Select
Rgts/Cal
from the menu bar.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......