A13914AF
9-37
Maintenance
Two-Month Maintenance
9
Replacement Procedure
1
To access this procedure, select:
•
Utils
from the menu bar
•
2 Maintenance
•
1 CTS Blade Replacement
.
To start the procedure, select
Continue
.
OR
To exit this screen, select
Cancel
.
2
The instrument prepares for blade and wick replacement, then the
Maintenance
pop-up
confirmation window appears.
CAUTION
The points of the Blade are very sharp and extend below the Wash Tower. Avoid
the bottom of the Wash Tower.
CAUTION
DO NOT select OK when the cover is removed. If you select OK the Cap Piercer
Carriage moves up and causes a pinch hazard.
3
Fully loosen the two captive Cap Piercer cover screws (1) at the base of the Cap Piercer cover.
Lift and remove the cover.
Figure 9.22
A014462P.EPS
1
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......