Table 12.10
Error Messages and Corrective Actions
ID
Class #
Pop-up
Event Log
Description
Corrective Action
0x01F80019
5
Cup positions were out
of order. Repeat
procedure starting with
position 1 which is
outermost.
CAU:
Align - Sample rack
positions are in reversed order
Message ID:
0x01F80019
Wrong order used for rack positions
while performing alignment.
•
Redo the alignment. Make sure the correct
positions are used (position 1 is on the outside of
the sample carousel).
0x01F90005
3
Reagent Carousel
Temperature Error.
Reagent Stability may
be compromised.
CAU:
Reagent Carousel
Temperature Error. Reagent
Stability may become
compromised
Message ID:
0x01F90005
Subsystem:
PWRMGR
Temperature:
Sensor indicates the air temperature
within the reagent carousel is out of
range (+2.0°C to +8.0°C).
Intermittent
message may be caused by:
•
recent loss of power
•
system reboot or
•
reagent door being open too long
No operator action is necessary.
Consistent
messages may indicate issues with the:
•
Thermister
•
Thermister cable
•
Power Control Board
•
Peltier Assembly
Contact Beckman Coulter Support Center for
additional information.
0x01F90084
4
Power Smart Module
Communication Error
CAU:
SMC Receive Error
Message ID:
0x01F90084
Subsystem:
PWRMGR
Msg Size:
number of bytes
A communication error has occurred
between the Power Distribution Board
(Smart Module 21) and the Instrument
Computer System (ICS).
•
Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
the system.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
0x01FA0003
4
DI water reservoir is not
filling at all
CAU:
DI water reservoir is not
filling at all
Message ID:
0x01FA0003
Whenever the DI water reservoir does
not fill within its given amount of time, it
will time out and flag that it has not filled.
•
Make sure DI water is turned on at the source.
•
Make sure the manual water inlet valve is
turned on.
•
Make sure the instrument has not rolled on top
of the water inlet lines.
•
and restart the system.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......