A13914AF
8-14
User Defined Reagents
Error Detection Limits
Mean deviation is a measure of noise in the read window.
Initial Rate High and Substrate Depletion
The initial rate parameter specifies maximum rate of absorbance change measured within the first
cycle after the last component (sample or reagent) is added. When this value is exceeded, results
are suppressed and flagged
Initial Rate High
(INIT RATE HI).
The delta absorbance specifies the maximum acceptable difference between the Reaction HIGH
ABS, measured just after sample or starter reagent addition, and the final absorbance measured at
the end of the reaction read window. When this value is exceeded, results are suppressed and
flagged
Substrate Depletion
(SUB DEPL).
Multipoint Spans
Multipoint span limits specify the allowable difference in absorbance or rate between respective
multipoint calibrator levels:
•
1 and 2
•
2 and 3
•
3 and 4
•
4 and 5
•
5 and 6
•
6 and 1
Values are only shown for the number of calibrators specified under "# of Calibrators". A change in
calibrator number resets the spans to zero (0.000).
The sign (positive or negative) of multipoint spans are dependent upon the direction of change
between two successive calibrators. For example, if the response for Cal 2 data < Cal 1 data, the
multipoint span must be negative.
Table 8.28
Allowable Entries for Reaction
Low Limit
-1.500 to 2.200
High Limit
-1.500 to 2.200
Mean Deviation
0.000 to 2.200
Table 8.29
Allowable Entries for Initial Rate and Delta Absorbance
Initial Rate
a
a. For downgoing reactions, the initial rate should be a negative number.
-99.999 to 99.999
Delta Abs
0.000 to 2.200
Table 8.30
Allowable Entries for Multipoint Span Limits
-1.500 to 2.200
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......