A13914AF
1-52
System Description
Cartridge Chemistry: Calibration Theory
Model #9
Math Model #9 is an extension to model #1, the four-parameter log-logit function.
The "c" is allowed to be 1 or -1.
If c = +1, then this is equivalent to model #1.
If c = -1, an alternative function is being used.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
Polynomial
Exponential
(PXP)
Quadratic
(POLY2)
Lorentz
Double Inflection
Model DP4
Table 1.18
Math Models for Non-Linear Chemistries
(Continued)
Type
Formula
E014476L.EPS
E014477L.EPS
E014478L.EPS
E011537L.EPS
R =
K
c1
+
K
c2
1 +
a
1
1
+
a
2
conc
conc