Maintenance
As-Needed/As-Required Maintenance
Figure 9.73
4
Connect the Probe Cleaner Tubing Assembly to the top of the probe assembly where the fitting
for the probe tubing was removed in Step 3. Tighten no more than finger tight.
5
Fill a 10 mL syringe with diluted wash solution. Connect the syringe to the other end of the
probe cleaning assembly.
6
Dispense all of the syringe contents through the probe into the collection cup. Repeat until
50 mL has been dispensed. Empty the collection cup as often as necessary.
7
Remove the cleaning assembly from the probe. Clean the probe tubing fitting with a cotton
swab wrapped in lintless tissue and moistened with deionized water. Dry the probe tube fitting
with a cotton swab wrapped in lintless tissue.
8
Reconnect the DI water line into the black fitting at the top of the probe.
9
Home the instrument.
•
Select
Instr Cmd
.
•
Select
1 Home
.
10
Prime the sample and reagent probes by performing the probe cleaning procedure listed in
Weekly Maintenance,
Clean Flow Cell, Cups and CC Probes/Mixers (Automated)
.
NOTE
Run controls to make sure that the cleaning completely removed the blockage.
A015947P.EPS
2
1
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......