1-28
System Description
Main Screen and Program Structure
Status Functions
The following Table depicts the status information available from the operator screens of the DxC
analyzer.
Table 1.13
Main Screen Status Functions
Status Indicator
Status Description
CTS
, appears in the blue bar at the top left side of the
Operator screens when the Closed Tube Sampling (CTS) option is installed. When
"CTS" appears, Closed Tube Sampling is enabled. When CTS appears within the
international "No" symbol (
), the feature has been installed, but it is not
enabled. When the indicator is absent, the CTS option is not installed. Enabling and
disabling Closed Tube Sampling is accomplished using the Setup procedures
described in the
Reserved Racks/Obstruct Detect
topic of CHAPTER 3,
System Setup
Options
.
IMPORTANT
CTS is an optional feature of the system, which allows the system to
pierce primary sample tubes. The operator is only offered the enable/disable
CTS option if a CTS assembly is installed, otherwise that selection is grayed out
on the
Setup
screen. The default for CTS sampling is "OFF."
Host
This indicator (2) in
, appears on the blue bar at the top right side of the
Operator screens. The indicator to the right of the label shows communication
activity between the instrument and the host computer. A blue bar indicates that
the host is sending information. A green bar indicates that the host is receiving
information.
Printing
The indicator appears at the middle of the operator screens. When the indicator
appears, you must reboot the DxC system to restart printing.
System Status
This indicator (3) appears on the left side of the
Main
screen in
just
above the
Menu Bar
(4). It shows the current state of the system: either
Running
,
Standby
or
Stopped
. When both the Modular Chemistry (MC) and the Cartridge
Chemistry (CC) functions have the same status, a single status indicator appears.
When the statuses of these components are different, the MC Status is shown on
the left, and the CC Status appears on the right. For example,
Running/Standby
would indicate that the MC side is
Running
, and the CC side is in
Standby
mode.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......