A13914AF
4-7
Reagent Load/Calibration
Reagent Load
4
CAUTION
To avoid contamination, do NOT touch the reagent straw.
NOTE
The No Foam container shown in
(2) is refilled by pouring a fresh bottle of No Foam
solution into the empty container on the instrument.
Refer to
and refill the No Foam container as follows:
•
Wipe dust from around cap (4).
•
Vent the No Foam container (2) by disconnecting the quick connector (3) at side of
container.
NOTE
The connector may be located on either side of the No Foam container.
•
Using the No Foam Container Cap Wrench, loosen the cap (4) from top of lid (1) by turning
the wrench counterclockwise.
NOTE
Make sure the wrench is flush with the No Foam cap.
•
Using the hand-held bar code reader, scan the bar code on new No Foam bottle.
•
Pour bottle contents into No Foam container (2).
•
Wipe off any spillage.
•
Replace cap (4) and hand tighten. Do not use the wrench to tighten the cap.
•
Reattach connector (3).
Figure 4.6
No Foam Container (DxC 800 shown)
1.
Lid
2.
No Foam Container
3.
Connector
4.
Cap
A016484P.PNG
2
3
4
1
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......