A13914AF
6-5
Sample Programming and Processing
Identify Samples
6
Sample ID
A Sample ID is a unique name used to identify the sample. Sample IDs may be specified with a
maximum of 15 alphanumeric characters. Sample ID is not a required field entry if a rack and
position assignment is made.
1
Select
Samples
from the menu bar.
2
From the
Program Samples
screen, select the
Sample ID
field to type the Sample ID.
Limitations
The instrument stores up to 10,000 unique Sample IDs. When this limit is exceeded, the system
overwrites the oldest sample program. Duplicate Sample IDs are not allowed. If Sample IDs are
reused by the laboratory, set the Maximum Sample Program Age feature to help alert you to
conflicts. Refer to CHAPTER 3,
System Setup Options
WARNING
If your LIS or normal workflow requires the reuse of sample IDs, the sample
programming should be cleared from the DxC at a time interval that is less than
the shortest time of sample ID reuse. The DxC over-write feature is not sufficient
for successfully reusing sample IDs. If this warning is not followed, results from
the new request are merged with tests from incomplete samples that previously
used that ID. In addition, patient demographics at the instrument are from the
previous use of the sample ID. Since demographics are not uploaded from the
instrument, it is possible for these merged results to be uploaded to the LIS which
would contain the current demographics.
Valid Characters for Entries
Valid and Invalid characters for entries are listed in the table below:
Table 6.2
Valid and Invalid Entries
Type of Entry
These are Valid ASCII
Character Codes...
Except for These Invalid Characters and
Character Codes...
Sample ID
33 thru 126
A
space
and $ , ; * ? [ ] \^ | &
Other
32 thru 175
ASCII Character Codes 44 and 127
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......