A13914AF
9-29
Maintenance
Monthly Maintenance
9
19
Select
OK
in the Maintenance pop-up confirmation window. The BUNm/UREAm cup is primed
with water.
20
Select
Close
when priming has completed to exit the
Result
pop-up window.
21
Select
Prime5,
to prime BUNm/UREAm 5 times with reagent. While the module is priming:
•
Observe the reagent line and reaction cup to confirm there are no leaks.
—
If any leaks are noticed correct the problem.
•
Observe the stir bar to verify that it is rotating properly.
—
Select
Close
when priming is complete.
22
Select
Drain
. When draining is complete, select
OK
.
23
When priming is complete, select
Close
to exit the
Result
pop-up window and select
Cancel
to
return to the Maintenance menu.
•
If the BUNm/UREAm electrode maintenance was performed, wait 15–20 minutes to allow
for thermal equilibration. Calibrate BUNm/UREAm.
•
To ensure there are no bubbles in the BUNm/UREAm cup, immediately prior to calibrating,
wrap the tip of an applicator stick with lintless tissue. Insert the end of the stick into the
bottom of the cup and carefully swab the electrode.
24
Proceed to
Calibrate Lamps and Sensor
Calibrate lamps and/or sensor following monthly cup maintenance, sensor replacement, or any
adjustment made to the module.
IMPORTANT
The system does not autoprime in
Maintenance
mode. After completing the maintenance
procedure, or if a procedure is aborted, exit the Maintenance menu to resume autoprime and return the
system to
Standby
.
1
Select
Utils
from the menu bar.
2
Select
2 Maintenance
.
OR
Type
2
in the
Option Number
field and press
(
Enter
)
.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......