0x01FA001A
4
10 psi Air Supply
Pressure Is High
CAU:
10 psi Air Pressure Is High
Message ID:
0x01FA001A
This error occurs when the transducer
line #129 is outputting a value into the
Hydro Board that is above the 10 psi
specification.
•
Check pressure levels and
the system.
•
Adjust pressure levels if necessary as follows:
—
Make sure the 25 psi (Main Pressure) gauge
on the hydro reads within 24 to28 psi.
—
Adjust the 25 psi (Main Pressure) regulator
as necessary.
—
When the 25 psi pressure is within range,
make sure the High Pressure reading on the
Hydro Status screen is 17 ± 1.
—
Adjust the low pressure system. The
pressure range for the low pressure system
is 8–11 psi. Set it at 10 psi.
•
Contact Beckman Coulter Support Center for
additional information.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......