aks-S4

Homecare bed with SMPS

Operating manual

Stand:  2015-10-15

Содержание aks-S4

Страница 1: ...aks S4 Homecare bed with SMPS Operating manual Stand 2015 10 15 ...

Страница 2: ..._____ 22 10 Accessories combination_ _______________________________________________ 23 11 Troubleshooting fault clearance__________________________________________ 24 12 Care cleaning_ _________________________________________________________ 25 13 Storage_ _______________________________________________________________ 26 14 Reuse__________________________________________________________________...

Страница 3: ...short circuit protection overcurrent and temperature shutdown minimal electromagnetic waves as a result no transformer humming is possible electricity saving because the power supply is in standby mode when the bed is not being operated safety due to 29 V instead of 230 V from mains adapter i e no 230 volts voltage in the power lead and at the homecare bed motorised height and back part adjustment...

Страница 4: ...re in a medical sector e g old people s homes and care homes rehabilitation facilities geriatric facilities The bed is suitable for adult patients and not for children The suitability of the homecare bed for the patient must be determined by the professional assessment of the health professional The homecare bed is only suitable for dry indoor areas The climatic conditions must be an ambient tempe...

Страница 5: ...anual safely for possible reference Include this operating manual in the case of change of ownership of the homecare bed Ensure the homecare bed and its accessories are in proper fault free condition before every use Note that the installation operation maintenance and repair of the homecare bed must only be performed by suitable qualified personnel You as the user and or owner must ensure e g by ...

Страница 6: ...that the side grill only provides protection against falling out if it is raised on both sides head and foot and if the lying area elements such as back thigh and lower leg part are in the horizontal position Only use original side grills as these have been approved by aks and ensure faultless and safe function Non original aks side grills can cause hazards When using the side grills check their s...

Страница 7: ...thorised dealer The aks S4 homecare bed is delivered in cartons with the following scope of delivery Table 01 Quantity Carton Contents 1 KA head support 1x head support Figure 02 with 6x Allen bolt pre assembled Figure 10 1 FA foot support 1x foot support Figure 01 with 6x Allen bolt Figure 10 pre assembled 2x extension tube connected to the frame Figure 06 1x carton with controller manual control...

Страница 8: ...aks S4 Figure 05 Figure 06 Figure 07 Figure 08 Figure 09 Figure 10 Page 8 z4200064_GA_Bett_S4_151015_EN ...

Страница 9: ...grills Castor with locking device Manual control unit Mains adapter Handle with strap Thigh and lower leg part Scissor lift Mirror panel Back part drive with controller Scissor lift drive Page 9 z4200064_GA_Bett_S4_151015_EN Product overview 5 ...

Страница 10: ...len bolts M8x16 with spring washer at the support mountings of the 2 head support Figure 13 Attach the head support with the support 3 mountings to the side round tubes of the scissor lift Let one end of the head support rest on the floor Figure 14 For fixing the head support screw the two 4 Allen bolts M8 x 16 with spring washer tightly into the support lugs on the side Figure 13 Remove all Allen...

Страница 11: ...ber see 8 arrow comes against the scissor lift drive You can set down the foot support in this position Figure 16 Hold the head and foot support in the middle and connect both supports together using the 9 extension tubes Figure 17 Screw all 8 Allen bolts tightly at the connection points of the lying area Figure 18 10 Remove the cable ties on the head and foot support 11 Temporarily connect the sc...

Страница 12: ... manual control unit to the controller and install the strain relief Figure 21 15 Now connect the plug connections for the drives according to Figure 22 16 H connection for manual control unit 1 connection for back part drive 2 connection for thigh part drive 3 connection for height adjustment drive in the scissor lift 4 blank plugs Z strain relief for manual control cable Note Strictly ensure the...

Страница 13: ...asten the connector cover strip to the controller Figure 23 17 Insert a mirror panel in the longitudinal tubes at the foot end of the lying area Figure 24 18 Fasten the mirror panel using 2 Allen bolts Attach the side grill bars to the side grill sliders and place the respective free end on the lying 19 area Figure 25 However only insert the second mirror panel at the head end so far into the long...

Страница 14: ... the mirror panel as far as the stop and fasten it using two Allen bolts Check the 23 side grill for function Insert the uplifter on either the left or right your choice in one of the head side receptacles 24 Figure 28 and if not already done attach the grab handle to the uplifter see section 8 3 Uplifter Insert the transport lock into the free uplifter socket see Chapter 25 9 Patient Transport Fo...

Страница 15: ...be laid so that it cannot be dragged driven over or endangered by moving parts during the operation of the homecare bed Also ensure that the mains adapter is always accessible in order to be able to disconnect the homecare bed from the mains power supply in an emergency Note The Medical Products Operator Regulations MPBetreibV are binding for the operating organisation user of medical products The...

Страница 16: ...0 Ensure that the castors are always braked unless you intend to transport the patient Side grills 8 2 The risk of the patient accidentally falling out of the homecare bed is reduced by the use of side grills In order to use the side grills pull these up slowly at the top side bar on the head and foot side until the side grill gliders engage To lower the side grills first raise the top side bar 1 ...

Страница 17: ...in particular the distances between bars and stays in relation to the physique of the bed user Ensure that the side grill height from the top side of the mattress is at least 220 mm If this is less than 220 mm use a side grill height extender Before every adjustment of the back part or thigh and lower leg part ensure that no limbs of the patient are between the side grills Uplifter 8 3 Position th...

Страница 18: ...so Figure 33 Patented angular adjustment of the lower leg part The aks S4 homecare bed is equipped with ratchet adjusters between the thigh and lower leg parts These enable the patient to adjust the lower leg part using the manual control unit without the otherwise required nursing personnel Figure 33 Perform the following steps for this Move the thigh part completely up In doing 1 so the lower le...

Страница 19: ...g so note that always only one function can be performed Otherwise the electrical system can disconnect due to overload If the manual control unit is not needed hang this on the manual control unit hook on the top side bar There is a rotary switch on the rear side of the manual control unit which is operated using the associatedkey Aswellasthe lock functiondescribed in Section 8 6 the manual contr...

Страница 20: ...e emergency lowering must only be performed by trained users Uncontrolled lowering of the back part can result in severe injuries for user and patient Practice the lowering under normal conditions so that you can lower the back part safely in an emergency Lock function 8 6 The manual control unit is equipped with a lock function i e the functions of the manual control unit can be blocked with the ...

Страница 21: ...cable and at the homecare bed Very low electrical and electromagnetic alter nating fields electrosmog during standstill of the drives 0 5 W in standby mode no constant transformer humming Improper handling can result in damage to the mains adapter and in hazards such as electric shock Do not pull on the cable of the mains adapter and do not drive over it Unload the back part before the emergency 1...

Страница 22: ...port in order to prevent the risk of driving over the 29 Volt power cable and the falling down of the mains adapter Ensure that none of the patient s limbs protrude from the bed and remove any equipment or objects attached to the bed Procedure The transport lock is located in the free uplifter holder Figure 40 The 29 Volt power cable is wrapped around the lying area and the mains adapter is insert...

Страница 23: ...ge cell mattresses Order No 39047 Side grill padding one side Order No 77551 Uplifter with grab handle Order No 77020 Other accessories on request Combination Other aks products can also be used with the homecare bed For example these include our aks antidecubitus systems or our aks patient hoists Only use mattresses whose dimensions correspond to the dimensions of the lying area also when using t...

Страница 24: ...s not function Lock function activated Unlock manual control unit Mains adapter not plugged in and 9 V block battery empty Plug in mains adapter and replace 9 V block battery 230 V room power socket has no voltage Check power socket fuse box No emergency lowering e g in the case of power failure 9 Volt block battery discharged Replace 9 Volt block battery Bed only moves very slowly Mains adapter n...

Страница 25: ...or cleaning materials containing ammonium chloride Only use dermatologically tested materials Do not use any basic and alkaline cleaning agents Do not use any aggressive cleaning materials e g solvents and hard brushes etc Only use alcohol free and chlorine free disinfectants and methods from the RKI Robert Koch Institut list or the disinfectants list of the VAH Verbund für Angewandte Hygiene e V ...

Страница 26: ...re beds is of course dependent on the type of use Frequent adjustment transport installation and cleaning reduce the service life as do improper handling irregular maintenance and exceeding the activation time of the drives or the safe working load Disposal 16 If the homecare bed has outlived its purpose the electrical components must be treated as waste electrical equipment according to the WEEE ...

Страница 27: ...roperly carried out maintenance work and technical modifications and additions attachments without the permission of the aks company result in voidance of the warranty and general product liability Declaration of conformity 18 The aks S4 homecare bed complies with the requirements of the EU Council Directive 93 42 EC Medical Products taking account of the amending Directive2007 47 EC The following...

Страница 28: ...dical product Check the 29 Volt power cable regularly at short intervals with respect to mechanical damage e g monthly and after every occurring mechanical load In order to rule out hazards in advance e g due to damaged insulation on the cables the mains adapter of the homecare bed must be unplugged from the mains power socket before the visual inspection In the case of damage to the supporting pa...

Страница 29: ...ise the back part and lower the side grill 1 Important Now unplug the mains adapter from the mains power socket Undo the fixing screw and remove the 2 controller from the back part drive Pay attention to the cable routing in doing so Using a Philips screwdriver undo the fixing 3 screws of the battery compartment cover and remove the cover Remove the battery from the battery 4 compartment and disco...

Страница 30: ...maged no wear sticker max 80 kg present 2 12 Transport lock for mains adapter present 3 Visual inspection of the electrical parts 3 1 29 V power cable and mains adapter SMPS not damaged 3 2 Strain relief kink protection applied and tight 3 3 Connection cable routing damage risk of getting caught 3 4 All enclosures of the electrical system OK O rings seals no cracks on cases fork heads and lift tub...

Страница 31: ...lower the back part by operating the manual control unit for a short time replace 9 V block battery if necessary Overall assessment Bed OK Date Company Inspector Signature YES NO If indications of defects damage arise during these tests the homecare bed must be disconnected fromthemainspowersupplyimmediatelyandmustnolongerbeoperated Obviousdefects damage must be checked assessed and repaired if ne...

Страница 32: ...te of manufacture Position head support foot support scissor lift back part drive thigh part drive scissor lift drive uplifter Rating plate Position uplifter Warning sign for removable side grill replaceable mattress Position foot support outside Warning sign maximum patient weight safe working load Position foot support inside ...

Страница 33: ...ion class II against electric shock Type B Attention Observe instructions for use only suitable for dry indoor areas WEEE marking the device must not be disposed of with the domestic waste This product complies with the basic requirements of the Medical Products Directive 93 42 EC ...

Страница 34: ...t temperature from 0 C to 40 C Relative humidity from 20 to 80 Air pressure from 700 hPa to 1060 hPa Normally composed atmospheric air Materials used Steel powder coated or galvanised Commercially available plastics POM ABS PP PVC PA6 6 Rubber Wood varnished or covered with decorative foil MDF plates covered with decorative foil Aluminium anodised Electrical Data Input 100 240 V 50 60 Hz 90 W Outp...

Страница 35: ...____________________________ Enter the year of manufacture of your homecare bed here Year of manufacture _____________________________ Enter the telephone number and name of the contact person of your authorised dealer here Name _____________________________ Telephone number _____________________________ Notes ...

Страница 36: ...aße 6 D 53842 Troisdorf 49 0 2241 9474 0 49 0 2241 9474 88 aks aks de www aks de Reprints whether in whole or in part are only allowed with the previous permission of the publisher All rights reserved We also reserve the right to make technical changes ...