12
0x02BC1300
4
Photometer Error (
error
message
), Command:
0xcommand
, Error
Number:
0xerror
number
CAU:
Photometer Error
Message ID:
0x02BC1300
Error Msg:
error message
Command:
0xcommand
Error number:
0xerror number
This error occurs when the photometer
fails to successfully read/analyze a
cuvette. There are several types of
Photometer Smart Module (PSM) errors.
If for any reason the instrument "Stops"
while the photometer is trying to read a
cuvette, the photometer will flag an
error. Check the Event Log for any
messages that might indicate a problem
around the same time stamp as the PSM
Error.
Check the Cuvette Center Alignment for
any cuvettes that exceed a 0±6 value.
This will flag a PSM Error.
•
the system.
•
Check Event Log for specific PSM subsystem
errors. Refer to
•
Contact Beckman Coulter Support Center for
additional information.
0x02BC2107
4
Failed to Initialize
Photometer
CAU:
Failed to Initialize PSM
Message ID:
0x02BC2107
The Photometer Smart Module (PSM)
failed to initialize during startup or
homing, the CC side is stopped.
•
Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
the system.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
0x032B0005
2
CC Obstruction
Detection Transducer
Failure
CAU:
CC Obstruction Detection
Transducer Failure
Message ID:
0x032B0005
Subsystem:
rasdSmc
This error usually occurs whenever the
obstruction transducer does not see fluid
during runtime. Often this comes from a
clogged degasser or lack of Degassed
water to the system. The obstruction
detection transducer is looking for a
certain range of pressure during its
primes before sample aspiration. If it
does not see the correct differential, it
will flag a transducer error.
•
Check Obstruction Detection Transducer for
obvious leaks or dripping from connections or
the device itself and CC Sample Probe.
•
Check for pinch in Water line (from syringe to to
top of probe).
•
Make sure the CC sample Syringe is properly
seated.
•
Make sure CC Sample Probe tubing is properly
connected at top of Probe. Refer to Step 3-4 of
the procedure.
•
Contact Beckman Coulter Support Center for
additional information.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......