A13914AF
9-111
Maintenance
As-Needed/As-Required Maintenance
9
5
Exit the Maintenance menu.
IMPORTANT
To resume autoprime and return the system to
Standby
, exit the Maintenance menu.
Clean the Work Surfaces
Clean all exposed work surfaces and covers that may have contacted sera or other biological fluids.
1
Remove reaction carousel cover (1) and sample carousel cover (2).
Figure 9.78
2
Scrub the covers with a good grade of disinfectant soap and a soft brush.
3
Rinse the covers in deionized water. Dry thoroughly using lintless tissue.
4
Reinstall the covers.
5
Wipe all exposed surfaces on the system with a 10% bleach
*
solution (one part 5.25% sodium
hypochlorite (PN A32319), combined with nine parts deionized water). The surfaces include:
sample rack area covers, pull down cover, compartment doors, keyboard, table surfaces and
sides of the console.
*
DO NOT use bleach that contains additives (for example, Ultra Bleach Advantage).
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......