A13
9
14
AF
12
-2
2
Tro
u
blesho
o
ting Cal
ibr
a
tio
n a
n
d Result
E
rro
rs
Comm
on Er
ro
r Me
ss
ages
and
C
o
rr
ec
tiv
e
Act
ion
s
0x04B20026
4
Sample tube did not
vent correctly. Remove
cap and reload tube.
Sample ID:
sample id
Rack:
sample rack id
Position:
sample
position
, MC Sample
Probe
CAU:
MC Sample Tube Excess
Vacuum
Message ID:
0x04B20026
Sample ID:
sample id
Rack ID:
sample rack id
Pos:
sample position
Probe detected excessive vacuum in
thick stopper closed tube on second
attempt. Sample is aborted.
•
Uncap tube and rerun.
•
If occurs consistently for many samples, check
cap piercer for broken blade. Refer to
CTS Blade/Wick (1-Blade Thick CTS Option)
or
Replace CTS Blade (1-Blade Narrow CTS Option)
to inspect the blade.
•
Perform MC ODC Alignment.
0x04B20027
4
Sample tube did not
vent correctly. Remove
cap and reload tube.
Sample ID:
sample id
Rack:
sample rack id
Position:
sample
position
, MC Sample
Probe
CAU:
MC Sample Tube Excess
Pressure
Message ID:
0x04B20027
Sample ID:
sample id
Rack ID:
sample rack id
Pos:
position
Probe detected excessive pressure in
thick stopper closed tube on second
attempt. Sample is aborted.
•
Uncap tube and rerun.
•
If occurs consistently for many samples, check
cap piercer for broken blade. Refer to
CTS Blade/Wick (1-Blade Thick CTS Option)
or
Replace CTS Blade (1-Blade Narrow CTS Option)
to inspect the blade.
•
Perform MC ODC Alignment.
0x04B3000F
4
ISE DAC Setting Failed
Chem Name:
chem id
CAU:
ISE DAC Setting Failed
Message ID:
0x04B3000F
Chem Name:
chem id
ISE DAC Setting Failed for the specified
chemistry.
•
Make sure the ISE module is in the down
position, the cover is in place, the reagent door
is closed.
•
Check for reagent volume in bottles or flow cell
leaks.
•
Check electrode connections.
0x04B30010
4
MC Reagent Level High
in Reaction Cup
Chem Name:
chem id
CAU:
Reagent too full
Message ID:
0x04B30010
Cup:
chem id
MC sample probe detected reagent in
the cup module above the expected
level.
•
Prime cup with reagent and make sure it drains.
Refer to
•
Failure to drain is either caused by low vacuum
or a valve failure.
•
Contact Beckman Coulter Support Center for
additional information.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......