A13914AF
6-11
Sample Programming and Processing
Additional Programming Information
6
Batch Programming
Batch Programming provides programming of multiple samples with the same chemistries, sample
type, dilution factor, and status. Any Panel, Comment, Demographics, Dilution Factor, Sample Type,
or Chemistry programmed will apply to all sample positions designated within the batch. This
information may be edited at any time on an individual position basis when the batch programming
is completed. Sample identification and description of each sample may be individually entered
later through the Sample Programming function. The Batch Mode is complete when End Batch is
requested or upon exiting the
Sample Program
screen.
Batch programming in rack and position mode can be performed using the following procedure.
1
Select
Samples
from the menu bar.
2
Select the tests, sample type, and any other data that is common for all samples in the batch,
including sample options information.
3
Select
Batch F4
. A list of available racks for batch programming appears. Racks may also be
cleared from this screen.
•
To exit batch program, select
Cancel
.
•
To clear racks, continue with Step 4.
•
To batch program, continue with Step 5.
4
To clear racks, type the specific rack ID in the
Rack
field and select
Clear
. A message appears,
and verifies that all programming for this rack will be deleted.
•
Select
OK
to delete.
OR
•
Select
Cancel
to keep current programming.
Either response will return the screen to the Batch programming screen.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......