A13
9
14
AF
12
-3
0
Tro
u
blesho
o
ting Cal
ibr
a
tio
n a
n
d Result
E
rro
rs
Comm
on Er
ro
r Me
ss
ages
and
C
o
rr
ec
tiv
e
Act
ion
s
0x05DC0204
4
MC Probe/Syringe Error
CAU:
Homing Error for MC
Probe/Syringe
Message ID:
0x05DC0204
Subsystem:
boot, homing
command
An error occurred while homing the MC
Probe or MC Sample Syringe drive.
Check the Event Log (classes 2 and 4) for
specific MC probe or syringe related
errors.
•
When the instrument goes to
Stopped
state,
check for possible interference with the motion
of the MC probe or MC sample syringe drive.
•
Make sure the ISE, MC cup and sample carousel
evaporation covers are installed properly.
•
Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
the system.
•
While the system Homes, check for possible
interference issues. If the error continues after
homing,
the instrument.
•
If the problem continues after the instrument
has been shutdown and rebooted, contact
Beckman Coulter Support Center for additional
information.
0x05DC0227
4
No Response From
Hydro Subsystem
CAU:
Homing Timeout for
Hydro Subsystem
Message ID:
0x05DC0227
Subsystem:
boot, homing
command
This error occurs when the hydro fails to
respond to commands or fails to attach.
This event can be generated by any
Hydro command sent by start.dat
including "AttachHydro",
"StartAutomonitor", "HydroInitStatus",
or "HydroWatchdogOn".
•
Check Event Log for specific subsystem errors.
•
the system.
•
Reboot. Refer to
.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
0x1391000E
Instrument stopped
communicating with
console
ROUTER:
Cable down
detected. (/mpcvan/src/
cauman/cau_router.c:1675)
Message ID:
0x1391000E
This Fatal Error will occur when a
network connection has been lost and
the computer console cannot
communicate with the instrument.
•
Make sure Instrument is on. Refer to
.
•
Check Ethernet cable connection at console,
microhub and instrument.
•
•
Contact Beckman Coulter Support Center for
additional information.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......