A13914AF
6-6
Sample Programming and Processing
Sample Programming and Processing
Create Sample IDs
When creating Sample IDs, use a format that distinctly differs from that used for Calibrator IDs. This
prevents the reporting of erroneous results due to calibrators being run as patient samples, or
patient samples run as calibrators.
Examples:
Calibrator Bar Code ID: MULTICAL
Sample Bar Code ID: 0000001
Sample Programming and Processing
Processing Bar Coded Samples
Generally, previous sample programming need not be cleared before placing a bar coded sample on
the instrument. If the lab reuses Sample IDs, previous programming must be cleared.
1
Load tubes in a rack with the bar code labels visible through the slot on the same side of the rack
as the rack bar code label.
2
For closed tube sampling systems:
•
Check the top of the validated closed tube for visible blood.
—
If blood is present,
·
Clean the cap with a cotton tipped applicator stick.
For non-CTS Systems or tubes not validated for closed tube sampling, remove the cap.
3
For a Routine Sample or if system is in
Standby
:
•
Priority samples should be placed on the autoloader first.
—
Place the rack in the autoloader with the rack bar code label to the right and
—
Press
RUN
.
For a STAT Sample:
•
If the system is running and there are other racks on the autoloader,
—
Press
PRIORITY
. The rack pusher moves back one space so the STAT rack can be placed
in front of the other racks.
—
Press
RUN
.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......