A13914AF
9-115
Maintenance
As-Needed/As-Required Maintenance
9
8
Prime the 10% bleach solution through the CC and MC subsystems as follows:
•
Select
Utils
from the menu bar,
•
Select
Prime
,
•
Select the
All CC Subsystems
check box,
•
Type
10
in the
Number of primes to repeat
field,
•
Select
MC F4
,
•
Select the following check boxes:
All cup chemistries
(ALBm, CREm, GLUCm, PHOSm, TPm, BUNm/UREAm)
DI water
Sample Delivery Subsystem
EIC wash
•
Select
Start Prime
. Allow all primes to complete.
9
Repeat Steps 1–8 three more times.
10
After the last prime is completed, allow the system to sit idle with 10% bleach solution for ten
minutes.
11
Remove both DI Water canisters and the Wash Concentrate bottle from the system. Use the
spanner wrench tool, if necessary. Empty out the bleach solution and rinse out the canisters
with DI water.
12
Install both canisters back onto the system using the spanner wrench tool, if necessary. Make
sure the large O-ring is in place and both canisters fit tightly.
13
Install the original Wash Concentrate bottle onto the system.
14
Refer to Step 7 and restart the hydropneumatic system.
15
Calibrate and run QC on all CC and MC chemistries.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......