A13914AF
9-118
Maintenance
As-Needed/As-Required Maintenance
Remove the Blade Assembly and Wick Clip Assembly
The Cap Piercer Assembly is now visible as shown in
. Remove the Blade Assembly and
Wick Clip Assembly as follows:
1
. Use a flat-head screwdriver to loosen the two Blade Holder Screws (1) from
Blade Holder (2).
CAUTION
The points of the Blade (3) are very sharp and extend below the Wash
Tower (4). To prevent possible injury or exposure, avoid the bottom of the Wash
Tower.
Figure 9.82
A007557P.EPS
1
2
4
3