A13914AF
9-21
Maintenance
Monthly Maintenance
9
16
Calibrate the CO
2
chemistry.
Adjust the Fluid Level of the Alkaline Buffer Damper Assembly
1
Visually inspect the damper to verify that the fluid level rests between the MIN and MAX lines,
approximately one-third to one-half full.
If the fluid level is below the MIN line, proceed to Step 2.
If fluid level is above the MAX line, go to Step 8.
2
Remove the alkaline damper (1) from the holding clip and place the damper in a horizontal
position.
Figure 9.10
NOTE
The damper should be positioned so that the output line #29 (2)
, leading from the
damper to the flow cell, is oriented on top.
3
Prime alkaline buffer five times or as many times as necessary to fill the damper one-half full.
Access alkaline buffer priming.
•
Select:
—
Utils
from the menu bar,
—
1 Prime
,
—
MC F4
,
—
ISE CO
2
Alkaline Buffer
.
A015938P.EPS
1
2
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......