A13914AF
9-76
Maintenance
As-Needed/As-Required Maintenance
Prepare the New CO
2
Measuring Electrode Membrane
1
Use tweezers to remove a new membrane (PN 661750) from the package in the Maintenance Kit.
Rinse the membrane thoroughly, on both sides, with deionized water. Gently dry the membrane
with lintless tissue.
Use the packaged membranes that are separated by pink spacers.
Figure 9.55
2
Place the membrane clamp (2) on a flat surface. Using tweezers, carefully center the membrane
(1) on top of the membrane clamp.
Figure 9.56