A13914AF
9-99
Maintenance
As-Needed/As-Required Maintenance
9
Replace the Sample and Reagent Probe
The Sample or Reagent Probe should be replaced when visual inspection or chemistry performance
indicates there is a compromised probe condition (scratches, pits or blockage). These instructions
cover installation of the replaceable style probe.
CAUTION
To prevent damage due to electrical static discharge (ESD), wear a wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
Replace the Old Probe
1
Press the
STOP
button on the instrument to stop all motors.
2
Using a 3/8 inch open end or adjustable wrench, loosen the nut (1) at top of probe.
Figure 9.74
3
Pull downward first on the probe, then up and to the side to remove the probe from the collar
wash.
4
Check to make sure the O-ring is on the nut. If the O-ring stayed inside the level sense bead,
bump it loose with a wooden applicator stick.
A007413P.EPS
1
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......