A13914AF
4-6
Reagent Load/Calibration
Reagent Load
Figure 4.5
DxC 600/800 Center Compartment
Modular chemistry reagent bottles are selected from the
Rgts/Cal
screen and replaced as necessary
according to volume. Reagent volume levels are automatically tracked by the system after the initial
volume is set. When a new reagent bottle is replaced the system assumes the bottle is 100% full. If
the new bottle is not 100% full, the volume may be adjusted.
Load/Unload Modular Chemistries (MC)
1
Select
Rgts/Cal
from the menu bar.
2
Select the
MC Rgts (Modular Chemistries)
.
3
Select
Load F1
.
4
Open the leftmost and/or center doors to access modular reagents. Refer to
or
.
5
Remove the reagent from the system. It is not necessary to scan the bar code of the reagent
bottle when removing it.
6
Place new bottle on the system.
1.
Wash Concentrate II
2.
No Foam solution
3.
CTS Auto-Gloss
A015922P.EPS
1
3
2
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......