A13914AF
9-65
Maintenance
As-Needed/As-Required Maintenance
9
Replace the Chloride Electrode
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
1
Disconnect the electrode cable from the CL connector.
2
Loosen the flow cell retaining screw (1) located at the top of the flow cell. This is a captive screw
and cannot be removed from the flow cell. Pull assembly forward and upward to remove.
IMPORTANT
Do not disconnect any of the reagent lines attached to the flow cell. Removal of the flow
cell is not necessary for replacing electrodes located on the right side of the flow cell.
Figure 9.43