A13914AF
12-13
Troubleshooting Calibration and Result Errors
Troubleshooting Result Errors
12
Troubleshooting Result Errors
Result Errors
Introduction
All reactions are checked against chemistry parameter flag limits (for example, absorbance limits,
reference ranges) to qualify the reagent, calibration, or sample. Checks are performed against the
final result and on interim reaction data for all samples, including calibrators. Any calibrator or
sample result or reaction step performing outside of any one or more limits will be flagged (message
or remark). A flagged sample value will not be reported.
Error Types and Description
Result Errors and Codes describe the errors or system conditions which affect chemistry results.
The conditions are described in
below.
Calibration Error Log
Error Flag
Magnitude
BLANK RATE
absolute delta
BLANK ABS
absolute delta
RXN RATE
absolute delta
RXN ABS
absolute delta
INIT ABS HI
actual value
INIT RATE HI
actual value
SPAN
absolute delta
All others
no magnitude
Table 12.7
Non-Linear Single Point and Multipoint Calibration Reports Field Definitions
(Continued)
Field
Description
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......