A13914AF
9-36
Maintenance
Two-Month Maintenance
Two-Month Maintenance
Replace CTS Blade/Wick (1-Blade Thick CTS Option)
Replace CTS Blade/Wick (1-Blade Thick CTS Option)
Refer to
. Replace the Cap Piercer blade and its CTS Auto-Gloss wick approximately every
two months. If the system runs >500 samples per day, replace the blade and wick more frequently.
The blade count records the number of caps pierced by the blade. Refer to
.
CAUTION
The Cap Piercer blade is very sharp and has been exposed to potentially
biohazardous fluids. To prevent injury or exposure, do not touch the points of the
blade and wear gloves.
Figure 9.21
Cap Piercer Blade with CTS Auto-Gloss Wick
Materials Required:
•
Medium Phillips screwdriver
•
Small Phillips screwdriver
•
Medium straight-edge screwdriver
1.
Blade Clamp
2.
Blade
3.
Points
4.
Wick Clip
5.
Wick (in clip housing)
A007367P.EPS
1
2
3
4
5
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......