A13914AF
9-120
Maintenance
As-Needed/As-Required Maintenance
4
Remove the old Wick Clip Assembly (1) from Wash Tower (2) by lifting up on both sides of Wick
Clip as shown in
. Make sure that the old Wick is retained within the Wick Clip. If the
wick is NOT in the Wick Clip, the Wick should be found within the Wash Tower area. In this case,
use tweezers to remove the Wick. Put the old Wick and Wick Clip into a biohazard waste
container.
CAUTION
Observe all laboratory practices or procedures that pertain to handling of
infectious and pathogenic materials.
Figure 9.84
A007559P.EPS