A13914AF
6-13
Sample Programming and Processing
Clear Samples
6
When to Clear Samples
Samples should be cleared as follows:
If sample programming is complete:
•
Clear sample programming.
If a sample ID must be reused:
•
Clear that sample ID first.
If a sample rack can not be programmed due to previous programming:
(This occurs when a sample ID does not match, or when a rack number shows previous information.)
•
Clear that programming.
Range of Sample IDs
The system retrieves a range of sample IDs in either Alphanumeric or Numeric form. The system
automatically selects one of the two, depending on the type of sample ID entered. These same
methods of retrieval are used for Load List, Clear Samples and Recall.
Sample IDs containing only
Numeric Data
:
•
Samples are listed in numeric order. For example 1, 2, 3, 4...
•
Include leading zeros in the
Range
field. For example, range 1 thru 1000 would not include
sample 0099, but range 0001 thru 1000 would.
Sample IDs containing only
Alpha
or
Alphanumeric Data
:
•
Samples are listed in dictionary order. For example A, AB, B...
•
Numbers come before letters. For example 1, 900, 1A, A...
•
Include all characters in the
To
field. For example, range A thru Z would not include ZEBRA, but
range A thru ZZZZZZZZZZZZZZ would.
IMPORTANT
Do not mix a pure numeric entry with an alphanumeric entry. For example, 0 through 1A
deletes all numeric.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......