A13914AF
9-81
Maintenance
As-Needed/As-Required Maintenance
9
4
Disconnect the solenoid valves, located on each side of the EIC, at the valve connector (1) as
shown below on DxC 800 Systems. The connector is located near the EIC ports where the ends
of the two solenoid valve cables meet. For DxC 600 Systems, trace the solenoid wiring to its
source and disconnect. Pull apart the valve connector to disconnect.
Figure 9.61
5
Carefully remove the Phillips screw and washer (1) (screw removed in figure) located on the top
of the EIC. Lift off the EIC (two sections with solenoid valves (2) attached) from the drip tray (3).
IMPORTANT
DO NOT REMOVE the two screws on the drip tray. Removal or repositioning of the drip tray
requires an MC sample probe alignment.
Figure 9.62