System Description
Cartridge Chemistry: Calibration Theory
1
Table 1.18
Math Models for Non-Linear Chemistries
Type
Formula
Model #1
Math Model #1 is the four-parameter log-logit function most commonly used with
reagents that use antibodies.
Sample values are determined using the calculated curve parameters and the math
model. Values may be calculated directly as this model can be solved for concentration.
Model #2
Math Model #2 is a five-parameter logit function.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
Model #3
Math Model #3 is a five-parameter exponential function.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
Models #4
through #7
These Models are reserved for future development.
Model #8
Math Model #8 is an alternative to model #2, the five-parameter logit function.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
E014421L.EPS
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......