A13914AF
10-9
System Status and Commands
Pause
10
Home
IMPORTANT
If the
STOP
button is pressed immediately after selecting
Home
:
•
Wait at least 10 seconds.
•
Press the
STOP
button again.
This makes sure that the system correctly reaches the
Stopped
state.
NOTE
If one side of the instrument (MC or CC) is
Stopped
and the other side is running, the Home command
operates on the side that is
Stopped.
The other side of the instrument continues to run.
1
Select
Instr Cmd
from the menu bar.
2
Select
1 Home
.
The instrument status changes to
Homing
. If the home completes successfully, the instrument
will be in
Standby
. Should an error occur, the status will be
Stopped
and the appropriate error
message appears.
Pause
Introduction
Pause
provides a way to prevent new tests from being started without wasting sample or reagent as
pressing STOP does. Those tests that already had reagent pipetted will run to completion, but no
new tests will be scheduled. This feature can pause the MC side, the CC side or both the MC and CC
sides of the instrument. The system will transition the paused side(s) of the instrument from
Running
to
Pausing
and eventually will reach
Standby
. To reach
Standby
status as soon as possible,
cuvette washing stops as soon as the last result is available. Cuvettes may be left dirty.
IMPORTANT
Do not request a
Pause
and leave the system sitting for a long period of time without washing
the cuvettes. If cuvettes are left sitting with reactant, evaporation will occur, leaving encrustation on the
sides of the cuvette that may require washing by hand.
Dirty cuvettes are washed automatically when the system status is
Running
or cuvettes can be washed
when the Maintenance option
Wash All Cuvettes
is selected.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......