A13914AF
7-7
Results Recall
Print Recalled Results
7
7
If
Yes
was selected in Step 6, type the password (if secured) and ID initials.
Select
OK
to return to the
Recall Results
dialog box with the result deleted.
OR
Select
Cancel
to return to the
Recall Results
dialog box and not delete the result.
NOTE
The deleted result with the initials of the operator are sent to an Event Log called Deleted Results.
8
Select
Prev F8
to view a previously shown sample.
OR
Select
Next F9
to view other samples if more than one sample was recalled.
OR
Select
Done F10
when completed.
Print Recalled Results
All recalled results may be printed.
1
Select
Results
from the menu bar.
2
To print using the default report format:
•
Go to Step 4.
To print using Patient Multi-Sample report format
•
Continue with Step 3.
3
If Patient Multi-Sample Report box is selected, input specific Patient ID. Select
Print F10
.
Select samples to be compiled into report.
All selected Patient Results are compiled into a single report.
4
From the
Recall Results By
dialog box, select the results to be viewed, as described previously in
this chapter.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......