Maintenance
Monthly Maintenance
9
BUNm/UREAm Electrode Maintenance (DxC 800 Only)
1
Drain the reagent from the MC cups.
•
Select:
—
Utils
from the menu bar,
—
2 Maintenance
,
—
4 Cup Maintenance
.
2
Select
BUNm/UREAm
to be drained in preparation for maintenance if performing BUNm/
UREAm maintenance.
Select
Drain
. A pop-up confirmation window indicates that the cup is drained and ready for
maintenance.
3
Loosen the two Phillips screws on the ISE module cover and remove the cover.
4
Loosen the two screws (1) on the MC cover and gently remove the cover.