Maintenance
As-Needed/As-Required Maintenance
Figure 9.75
CC and MC Sample Probe
Figure 9.76
Reagent Probe A and B
Confirm Instrument Performance
When the probe installation is complete, run the controls to confirm proper instrument
performance. If the controls are out but there are no visible leaks, recalibrate all chemistries and
rerun controls.
1.
Sample Probe
2.
Collar Wash
3.
Waste Port
4.
Wash Port
5.
Probe tip is even to the bottom
edge of the wash port
1.
Reagent Probe
2.
Collar Wash
3.
Waste Port
4.
Wash Port
5.
Probe tip is even to the middle of
the wash port
A007375L.EPS
1
2
3
4
5
A007376L.EPS
1
2
3
4
5