CHAPTER 9
Maintenance
Overview
Introduction
Maintenance on the UniCel DxC System should be performed by trained personnel to insure
continued, optimal performance and to reduce unnecessary service calls. Minimum training
requires that the operator read the Maintenance procedures.
The Scheduled Maintenance Log feature allows the system to keep track of what maintenance is
needed, when it is needed, and who performed the procedure. It also provides the capabilities of
alerting the operator to the maintenance that is due, and to the maintenance procedures that have
been performed and reviewed.
Basic Laboratory Practice
The maintenance frequencies established in this manual are based on processing approximately 400
samples/day.
The number of samples processed by the system, combined with the nature of the samples, may
require the stated maintenance frequencies to be adjusted in some laboratories.
A start-up prime of the various subsystems may be necessary if the system has been idle for more
than approximately 8 hours. Refer to
for priming procedures. Reviewing
reagent status prior to running the instrument is standard laboratory practice.
As with any laboratory equipment, the exterior of the DxC system should be kept clean. Any spills
should be cleaned up promptly. Spills that may be caused by leaking reagent lines or fittings should
be investigated. Refer to
As-Needed/As-Required Maintenance
in this chapter for cleaning
procedures.
Certain procedures in this manual require the use of a grounded wrist strap while the procedure is
being performed. The strap portion should be placed around the wrist. The wire portion can be
connected to the system in one of the following ways:
•
Alligator Clipped: Wrist straps with an "alligator" clip should be clamped onto the metal frame
of the system. To access, open either the left or center door. If desired, the doors may be closed
after the clip is attached to the metal frame.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......