A13914AF
9-66
Maintenance
As-Needed/As-Required Maintenance
3
Locate the Chloride electrode (1). Remove electrode by turning the electrode retaining nut.
Figure 9.44
4
Remove the retainer nut from the old electrode. Remove the quad-ring (1) from the tip of the
old electrode. Inspect the electrode port if the quad-ring is not on the electrode. Discard the
quad-ring.
Figure 9.45
5
Unscrew the old tip from the electrode assembly. Discard the O-ring and the old electrode tip.
6
Install a new O-ring on chloride electrode tip and screw the tip onto electrode body.
7
Completely dry the electrode port with lint-free tissue.
A015946P.EPS
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......