12
0x05DC01FB
4
CC Reagent Carousel
Motion Error
CAU:
Homing Error for CC
Reagent Carousel
Message ID:
0x05DC01FB
Subsystem:
boot
,
homing
command
This error occurs when there is any
homing issue with the Reagent Carousel
Subsystem. This includes the following:
•
Upper/Lower reagent decks/motors.
•
Both decks flag sensors.
•
The magnetic door sensor.
•
If reagent cartridges have been loaded recently,
make sure cartridges are fully seated and
refrigerator door is latched.
•
the system.
•
Check Event Log for specific RGTWHL
subsystem errors.
•
Contact Beckman Coulter Support Center for
additional information.
0x05DC01FD
2
CC Sample Mixer Error
CAU:
Homing Error: $s
Message ID:
0x05DC01FD
This motion error is contained within the
Sample Mixer Crane and will only occur
during a home. This can be either a
vertical or rotational error.
•
Make sure mixer has the necessary room to
move.
•
Make sure the wiring and cabling are secured
properly.
•
Make sure mixer paddle is seated firmly in its
housing.
•
If mixer appears out of position, perform rotary
or vertical alignment.
•
and restart the system.
•
If motion error continues, contact Beckman
Coulter Support Center for additional
information.
0x05DC01FF
4
Sample Autoloader
Error
CAU:
Homing Error: $s
Message ID:
0x05DC01FF
This error encompasses all homing
motions in the Sample Rack Handling
System (SRHS) and occurs whenever
there is a homing issue during an
initial
boot
. This includes the On Load
Assembly, Shuttle, Sample Wheel, and
the Lifter. If the devices do not find their
home position in time or they jam, the
error will be flagged.
•
Check for any racks that may have jammed in
the Sample Rack Handling System or if any of
the devices have jammed together.
•
Check for other obstructions if no rack problem
(covers, stoppers, etc.)
•
the system.
•
If error continues after
, print screen and
check event log for specific sub-system homing
error. Contact Beckman Coulter Support Center
for additional information.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......