A13914AF
12-16
Troubleshooting Calibration and Result Errors
Error Codes and Definitions
RACK CONFLICT
HDLC/IBCT/HbA1c sample in wrong rack,
test cancelled
RESULT ERROR
Result error
Results Suppressed
AG EXCESS
AX
Antigen excess
Results Suppressed
BL ABS HI
BH
Blank absorbance high
Results Suppressed
BL ABS LO
BL
Blank absorbance low
Results Suppressed
BL MAX DEV
BO
Blank outlier (maximum deviation)
Results Suppressed
BL MEAN DEV
BN
Blank noise (mean deviation)
Results Suppressed
BL RATE HI
SH
Blank rate high
Results Suppressed
BL RATE LO
SL
Blank rate low
Results Suppressed
CAL REF DRIFT
EC
Excessive reference drift – sample to
calibration
Results Suppressed
INIT ABS HI
AH
Initial absorbance too high
Results Suppressed
INIT ABS LO
AL
Initial absorbance too low
Results Suppressed
INIT RATE HI
IR
Initial rate too high
Results Suppressed
INIT RATE LO
IL
CC chemistries, initial rate too low
MC chemistries, a stirrer failure
Results Suppressed
INIT ADC HI
HI
Initial ADC error high
Results Suppressed
INIT ADC LO
LI
Initial ADC error low
Results Suppressed
INIT COND HI
CH
Initial conductance high
Results Suppressed
INIT COND LO
CL
Initial conductance low
Results Suppressed
K REQ
IK
Bad K value
Results Suppressed
NA/K REQ
IN
Bad NA value
Results Suppressed
OIR HI
DH
Out of instrument range high
Results Suppressed
OIR LO
DL
Out of instrument range low
Results Suppressed
ORDAC HI
OH
Out of instrument range ORDAC high
Results Suppressed
ORDAC LO
OL
Out of instrument range ORDAC low
Results Suppressed
URDAC HI
GH
Under range URDAC high
Results Suppressed
URDAC LO
GL
Under range URDAC low
Results Suppressed
ORR HI
UH
Out of reportable range high
Results Suppressed
ORR LO
UL
Out of reportable range low
Results Suppressed
ORR O HI
UO
Out of ORDAC reportable range high
Results Suppressed
OVERFLOW
OF
Number overflow error
Results Suppressed
REF NOISE
DR
Reference signal noise (ISE only)
Results Suppressed
RX ABS HI
HR
Reaction absorbance high.
Results Suppressed
RX ABS LO
LR
Reaction absorbance low
Table 12.9
Result Errors and Codes
(Continued)
Result Displayed on Report
Printed in
Remarks Area
Host
Code
Inst.
Code
Description
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......