A13914AF
9-45
Maintenance
Three-Month Maintenance
9
3
The syringe rods (1) are now positioned for placement.
Figure 9.29
4
Unscrew the barrel of the reagent syringe (left back wall) or barrels of two sample syringes
(back wall above modular cups) to release each syringe from the Luer-lock fitting. Refer to
5
Unscrew the round plunger end (1).
Figure 9.30
A015944P.EPS
1
A011874P.EPS
1