A13914AF
9-127
Maintenance
As-Needed/As-Required Maintenance
9
5
Install a new ISE drain pump tube:
•
Connect the tube to the tube fitting and route it through the pinch clamp on one side of the
pump.
•
Rotate the pump roller mechanism approiximately half a turn while feeding the tube under
the pump rollers.
•
Route the tube through the second pinch clamp and attach it to the tube fitting on the other
side of the pump.
•
Close the plastic cover.
6
Verify that the tube is properly positioned under the pump rollers and is not twisted.
7
Select
Close
to exit the maintenance procedure and return the system to
Standby
.
8
Select
Exit F10
9
Select
1 Prime
.
10
Select:
•
All Hydropneumatic Subsystem
from the Prime dialog box.
•
Type
5
in the
Number of primes to repeat
field.
•
Start Prime
.
11
As the ISE primes, check for leaks or bubbles from the tube or fittings, ISE waste draining, and
the tube seated properly.
If you see a problem, select
Stop Prime
. Check the fitting connections or replace a damaged tube,
starting again at Step 1 above.
12
Prime the ISE 20 times and watch for bubbles or leaks.
13
Select
Exit F10
to return to the Utilities screen.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......