A13914AF
9-34
Maintenance
Monthly Maintenance
4
Refer to
.
•
Fully loosen the one captive blade screw (2).
•
Fully loosen the two captive blade holder screws (1).
•
Remove the blade holder.
Figure 9.20
5
Hold the blade at the top of its shaft. Do NOT touch the points of the blade.
6
Lift up the blade to remove it.
7
Discard the old blade into a biohazard sharps container.
NOTE
The points of the blade are very sharp.
8
Remove a new blade from its container.
9
Hold the blade holder with the blade hole on the top. Note that the blade hole is keyed and the
blade has a slot.
Install the blade into the blade holder as follows:
•
Rotate the blade, if necessary, until it falls into place.
•
Finger tighten the captive blade screw.
•
Remove the protection from the points of the blade.
•
Do NOT touch the points of the blade.
A014
8
50P.EPS
1
2
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......