A13914AF
6-10
Sample Programming and Processing
Additional Programming Information
Additional Programming Information
Chemistry Identifiers
Color is used to indicate chemistry information.
•
The
chemistry name
appears in
red
when the chemistry is not runnable. There are two possible
reasons why a chemistry can not be run:
1.
The reagent is not loaded or
2.
A loaded chemistry has a reagent or calibration status that does not permit it to run.
Refer to the
Reagent/Calibration Status
screen for the reason the cartridge can not be run.
•
A
yellow indicator
appears when the chemistry is set for manual ORDAC.
•
A
blue indicator
bar turns from gray to blue when a chemistry is selected.
Enter a Dilution Factor
A dilution factor may be entered to be applied to the results of a specified sample by using
Options F3
. The dilution factor represents an off-line dilution prepared by the operator. The default
dilution factor is 1X. Each result for the sample are multiplied by the factor. Any final result
generated by the system is multiplied by the factor.
NOTE
The dilution factor will remain as part of the sample programming and must be removed if performing
additional testing of the undiluted sample.
Programming a Manual ORDAC
For samples with analytes of known concentration that exceed the analytic range, the manual
ORDAC feature may be selected. ORDAC compensates for extremely high concentrations by taking
either a smaller sample volume or diluting the sample online. Selecting manual ORDAC in Sample
Programming sends the request to run the test in ORDAC the first time it is run, unlike chemistries
selected for Automatic ORDAC (which are run undiluted first and then rerun using ORDAC when
recovery is out of range).
ORDAC is NOT available for uric acid that is run on urine samples.
NOTE
All ORDAC results are designated in the instrument code section on the appropriate results report.
IMPORTANT
Results which include the "less than" (<) sign are also multiplied. For example, a CRP result
with a sample dilution of 4, that should be reported out as "<0.5", is reported out as "<2.0".
Programming Samples for Serum Index
Specified samples can be programmed for serum index by using
Options F3
.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......