A13914AF
9-69
Maintenance
As-Needed/As-Required Maintenance
9
4
Lower the electrode face (2) into the soaking solution until only the tip is covered by the soaking
solution (1).
Figure 9.47
Removal and Installation of the Sodium Electrode
For access to the Sodium Reference electrode, instrument covers must be removed. To remove the
ISE Module cover, follow the steps below. The instrument status must be
Standby
to perform this
procedure
CAUTION
To prevent damage due to electrical static discharge (ESD), wear a wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter..
1
Loosen the two Phillips screws on the ISE Module cover and remove the cover.
2
Release the locking pin on the ISE Module and lift the module until it clicks into a raised, locked
position.
3
Drain the flow cell.
Select:
•
Utils
from the menu bar,
•
2 Maintenance
,
•
3 ISE Service
.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......