A13
9
14
AF
12
-3
4
Tro
u
blesho
o
ting Cal
ibr
a
tio
n a
n
d Result
E
rro
rs
Comm
on Er
ro
r Me
ss
ages
and
C
o
rr
ec
tiv
e
Act
ion
s
0x139F0069
9
Full CTS Tracking is now
available.
When you move a tube
to a different
instrument, it is not
necessary to remove its
cap.
STM:
TS-CTS Tracking System
reconnect detected
All of the systems that we expect to be
networked together are connected (it
has recovered from a disconnect state).
•
No operator action is necessary. Databases
automatically synchronize and caps can be left
in place.
0x13A00004
10
Requested Rack Not
Loaded
Rack:
rack ID
WLM:
Rack
rack ID
cannot be
loaded
(
FileName
:
LineNumber
)
The sample rack subsystem was unable
to load a rack, usually due to a motion
error. The rack will be removed and
processing will continue.
•
Correct the accompanying motion error and
reload the rack.
0x13A0001C
10
Non-programmed
samples on rack.
WLM:
Unprogrammed samps,
Rack:
rack ID
[p:
PositionSample ID
]
[p:
PositionSample ID
]
[p:
PositionSample ID
]
[p:
PositionSample ID
]
Rack has been loaded with a sample
which has not been programmed.
•
Make sure the complete sample bar code is
readable through the window in the rack.
•
Check communication with LIS.
•
Manually program if necessary.
0x13A00022
10
The QC Database is
xx
% full. Archiving is
suggested.
WLM:
QC DB is
xx
% percent
full
QC data base is becoming full. When the
database becomes full, the oldest
results will be overwritten. To avoid this,
archive your QC data.
•
data so it will not be overwritten.
0x13A00023
10
QC Database is full. The
oldest result has been
overwritten.
WLM:
QC DB is full. The oldest
result for
control:
control name
/
Chem:
chem name
has been
overwritten.
QC database is full. The oldest result for
specified control and chemistry has
been overwritten. Archiving is
suggested.
•
so that additional data will not be
overwritten.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......