background image

Labnet Spectrafuge™ 6C 

Instruction Manual

Catalog Numbers: 

C0060
C0060-230V

This manual is available in additional languages at 

www.labnetlink.com

.

Содержание Spectrafuge 6C

Страница 1: ...Labnet Spectrafuge 6C Instruction Manual Catalog Numbers C0060 C0060 230V This manual is available in additional languages at www labnetlink com...

Страница 2: ...Removing and Installing the Angle Rotor 3 6 3 Overloading the Rotor 5 7 0 Operation 5 7 1 Closing the Lid 5 7 2 Lid Release 5 7 3 Lid Lock 5 7 4 Speed Selection 5 7 5 Selection of Operating Time and M...

Страница 3: ...iquid spillage may harm unit Never put hands in the rotor area unless the rotor is completely stopped Never move the centrifuge while the rotor is spinning Never use solvents or flammables near this o...

Страница 4: ...1 each 5 0 Installation The centrifuge should be installed on a rigid even surface such as a stable laboratory bench cabinet etc To guarantee sufficient ventilation ensure that the centrifuge has at...

Страница 5: ...ft adapter see figure 1 Place the rotor on the motor shaft figure 1 and over the motor shaft adapter see figure 1 and 2 NOTE Figure 1 and 2 are located on the following page When loading the rotor ref...

Страница 6: ...4 LABNET SPECTRAFUGE 6C A Motor B Motor shaft adapter C Motor Shaft Figure 1 Chamber and motor shaft Figure 2 Bottom of angle rotor Figure 3 Loading the rotor...

Страница 7: ...mechanical damage Never centrifuge strongly corrosive materials that may damage the rotors or accessories 7 1 Closing the Lid After the rotor has been properly secured and loaded close the centrifuge...

Страница 8: ...and rotor accessories clean All parts should be wiped down periodically with a soft cloth For more thorough cleaning use a neutral cleaning agent pH between 6 and 8 applied with a soft cloth Excessive...

Страница 9: ...rvice Centrifuge is not receiving power See Centrifuge will not start Centrifuge cannot be started although power is on Lid not closed correctly Close lid correctly No speed or time has been selected...

Страница 10: ...ues The centrifuging radius of the 15 mL rotor is 8 5 cm for round bottom tubes and 8 2 cm for conical bottom tubes See Section 6 for the correct radius when using adapters and smaller tubes The chart...

Страница 11: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...

Страница 12: ...r local Corning sales office www labnetlink com For a listing of trademarks visit www corning com clstrademarks All other trademarks are the property of their respective owners 2021 Corning Incorporat...

Отзывы: